Sartorius CH3G Instructions d'exploitation Page 63

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 112
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 62
Range 1, Range 2, Range 3
Here you can enter the limits for each
of the weighing ranges. When a limit
is exceeded, the accuracy changes.
The following applies when entering
range limits:
Range 1 < range 2 < range 3
< maximum capacity.
Thus, you can divide the weighing
capacity into four ranges.
The resolution changes in intervals
of 1, 2, 5, 10, 20, 50, etc. The lowest
resolution is equal to the specified
lowest scale interval d. Set unused
ranges to 0.
Maximum Capacity (Max. Cap.)
The maximum capacity is the maximum
load that may be placed on the
weighing instrument. If a heavier load
is placed on the platform, the display
shows
H.
The scale intervals are calculated from
the maximum capacity and the lowest
scale interval d. In legal metrology,
the number of intervals must not
exceed 3000 e; for multi-interval
scales, the limits is 3000 e per range.
In Standard operation, as opposed
to legal metrology, you can define
a “SuperRange” weighing instrument
with more intervals. For example, a
maximum capacity of 30,000 kg with
a lowest scale interval d of 0.001 yields
30,000 scale intervals. These parame-
ters, however, may be influenced by
physical restrictions.
Available Units
With this function, you can make
particular weight units (Weight unit x;
x=1, 2) inaccessible during weighing.
Available units are indicated by a circle
(o) on the display (more than one
can be selected). To enable or disable
a unit, select the unit by pressing
the Q or q soft key, and then press
the l soft key (toggle function).
Calibration/Adjustment Unit
This setting defines the weight
unit with which calibration must be
performed. The calibration unit remains
the same, even when a different weight
unit is used during normal weighing
operation. The selected weight unit is
marked by a circle (o). To change the
weight unit, select the desired unit by
pressing the Q or q soft key and press
the l soft key to confirm.
63
Save Parameters
Select “Yes” by pressing the Q soft key
and press l to confirm; the A/D con-
verter configuration settings are stored.
The device software is reset, and the
scale returns to the normal weighing
mode. The Service mode is deactivated.
To exit the menu without saving
configuration changes, press the
o soft key.
Once these parameters have been
configured, the A/D converter in
conjunction with the load cell(s) is
defined as a weighing instrument.
The A/D converter, in conjunction with
the weighing platform, can now be used
like any standard weighing platform.
In addition, the weight unit must
be defined and the weighing platform
adjusted (calibration, adjustment
and linearization must be performed).
For a detailed description of these
procedures, see the chapter entitled
“Calibration and Adjustment.”
Adjust without Weights
Open this submenu from the
“Calibration/Adjustment” menu
(refer to the menu tree diagram
on the previous pages).
Open the “Input parameters”
submenu to enter the parameters for
“Nominal load” (in kg; specification of
the strain-gauge beam), “Resolution”
(in kg) and “Sensitivity” (in mV/V;
under “Sensitivity 1”). These values
are converted to internal quantities.
Once the A/D converter configuration
data has been stored (by selecting the
“Save parameters” menu item), these
parameters can no longer be read. For
weighing platforms with multiple load
cells, you can either enter sensitivity
of the other load cells individually
(“Sensitivity 2” through “Sensitivity 4”
for load cells 2 through 4), or enter
the mean value of all sensitivities under
“Sensitivity 1” (and enter “0” for
“Sensitivity 2” through “Sensitivity 4”).
Save the values entered for these
parameters by selecting the “Save
parameters” menu item.
Geographical Data
This is a submenu of the “Calibration/
Adjustment” menu (see the menu tree
diagram on the previous pages) and is
described in detail in a separate section,
starting on page 79.
! Note:
After configuring the A/D converter
and adjusting the weighing instrument
(calibration, adjustment and lineariza-
tion), return the menu access switch (17)
to the “closed” position.
Once the A/D converter configuration
has been locked (menu access switch
closed), the indicator can no longer be
used to influence weighing results.
The scope of functions available in the
weighing instrument is defined by the
A/D converter. Weighing functions that
can be activated include:
Read weight values
Tare the scale
Calibration/adjustment
Read tare value
Save/delete tare input
Zero the scale
Vue de la page 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 111 112

Commentaires sur ces manuels

Pas de commentaire