Care and Maintenance
Cleaning the Interior
of the Weighing Platform
Important note:
This does not apply to CH model
platforms.
$ If dirt gets inside the weighing platform,
the load plate has to be removed to clean
the weighing platform interior. Particular
caution is advised when working with
models 1000 + 1000 mm or larger.
$ Blow out the interior with pressurized
air or rinse with a weak water jet
(max. 60°C).
Make sure that no dirt builds up in the
gap between the load receptor and the
fastening plate to avoid compromising
the overload protection.
Cleaning Stainless Steel Surfaces
Clean all stainless steel parts regularly.
We recommend removing the stainless
steel load plate and cleaning it separately,
outside the hazardous area. Use a damp
cloth or sponge to clean any stainless steel
parts on the scale. You can use any com-
mercially available household cleaning
agent that is suitable for use on stainless
steel. Clean stainless steel surfaces by
wiping them down.
Then clean the load plate thoroughly,
making sure to remove all residues.
Use a damp cloth or sponge to wipe down
any stainless steel parts on the weighing
platform again and allow the equipment
to dry. If desired, you can apply oil to the
cleaned surfaces as additional protection.
!Do not use stainless steel cleaning agents
that contain soda lye, ethanoic acid,
hydrochloric acid, sulphuric acid or citric
acid. The use of scrubbing sponges made
with steel wool is not permitted.
Solvents are permitted for use only on
stainless steel parts.
Corrosive Environment
$ Remove all traces of corrosive substances
from the weighing platform on a regular
basis.
Replacing the Dust Cover
> Replace damaged dust covers.
§ Place the new dust cover on the
indicator and press down on the front
and back along the edges until the cover
is firmly seated.
Safety Inspection
Safe operation of the device is no
longer ensured when:
there is visible damage to the device
or power cord
the built-in power supply no longer
functions properly
the device has been stored for a
relatively long period under unfavorable
conditions (e.g., extreme moisture)
the equipment has been subjected to
rough handling during shipment.
If there is any indication that safe
operation of the device is no longer
warranted:
§ Disconnect from AC power (unplug the
equipment from the wall outlet (mains
supply))
> Lock the equipment in a secure place
to ensure that it cannot be used for the
time being
§ Notify your nearest Sartorius Service
Center or the International Technical
Support Unit based in Goettingen,
Germany.
Maintenance and repair work may be
performed only by authorized Sartorius
service technicians who:
have access to the required service and
maintenance manuals, and
have attended the relevant service
training courses.
!The seals affixed to this equipment
indicate that the equipment may be
opened only by authorized service
technicians, to ensure safe and trouble-
free operation of the equipment and to
maintain the conditions for warranty
coverage. If the verification seals are
damaged, the equipment must be
reverified.
Service
Regular servicing by a Sartorius
technician will extend the service life of
your Combics scale and ensure its
continued weighing accuracy. Sartorius
can offer you service contracts, with your
choice of regular maintenance intervals
ranging from 1 month to 2 years.
The optimum length of the service
interval depends on the operating
conditions at the place of installation
and on your requirements.
Cleaning
!Unplug the equipment from the AC
adapter or power supply before
cleaning.
$ Observe your company’s internal
regulations as well as standard industry
guidelines with regard to cleaning
intervals and cleaning agents.
$ If the scale is located in a dry room,
wipe down the weighing platform with
a damp cloth. You can use common
household cleaning agents. Follow the
cleaning agent manufacturer’s
instructions.
!Never use concentrated acids, alkali
solutions, solvents or pure alcohol to
clean the equipment.
$ If the scale is located in a wet room,
clean the weighing platform from above
with a weak water jet (max. 60°C).
!Do not use high-pressure cleaning
equipment to clean the weighing
platform.
> If the water that you use to clean the
weighing platform is too hot or too
cold, the difference in temperature
between the water and the weighing
platform can cause condensation within
the weighing platform. This
condensation can lead to malfunctions
in the equipment.
$ If the weighing platform is installed in a
pit, make sure that no dirt accumulates
between the edge of the pit and the
weighing platform. This will help
prevent measuring errors.
$ Regularly remove all dirt from the floor
of the pit.