Sartorius YDP 03-0CE Instructions d'exploitation Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 117
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
17
§ To record consecutive weighing operations, follow the
steps mentioned above. Then:
§ Press the key to obtain a printout of the statistics
> The values, as shown on the left, are automatically
computed and printed out
> Once the data has been printed out, the statistics memory
will be cleared
Practical Example
Suppose you wish to weigh tablets and have their weights
statistically evaluated. Let’s assume that your weighing
instrument is used in legal metrology.
You need to make the following settings:
Generate an ISO/GLP-compliant printout = 1 (-a1-)
Automatic taring after each data transfer = 1 (-a2-)
Statistics program = 1 (-a3-)
Printout after each data transfer = 1 (-a4-)
Index number for each data printout
(numbering function) = 1 (-p4-)
Automatic data transfer = 0 (-p6-)
$ In addition, set the menu code for an ISO/GLP printout on
your weighing instrument:
refer to the “Installation and Operating Instructions” for
this instrument.
§ Press the key to clear the statistics memory prior to
initial data transfer
> Any previous values remaining will be evaluated first,
then you will obtain a printout of the statistics.
§ Place the first tablet on the weighing pan
§ Activate data transfer by pressing the key
> When the weighing instrument has stabilized, this value
will be transferred and read into the statistics memory
as the reference weight
123456789012345678901234
n 12
x 99.8 g
s 3.55 g
srel 3.57 %
x 1203.4 g
min 96.0 g
max 103.2 g
Diff 7.2 g
12-DEC-95 16:02:11
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 116 117